Ramesh, P. J. and Basavaiah, K. and Vinay, K. B. and Xavier, C. M. (2012) Development and validation of RP-HPLC method for the determination of Ganciclovir in Bulk Drug and in Formulations. ISRN Chromatography, 2012.
Text (Full Text)
Development and Validation of.pdf - Published Version Restricted to Registered users only Download (2MB) | Request a copy |
Abstract
A simple, rapid, accurate, and precise gradient reversed-phase HPLC (RP-HPLC) method has been developed for the determination of ganciclovir (GNC) in pharmaceuticals. Chromatographic separation was carried out on inertsil ODS C18 (4.6 mm i . d × 2 5 0 mm, 5.0 μm) LC column using ammonium acetate buffer, sodium salt of hexane sulfonic acid as ion-pairing reagent in 1000 mL water, and acetonitrile (90 : 10) (v/v) as mobile phase at a flow rate of 1.0 mL m i n − 1 and with UV detection at 245 nm at column temperature (30°C). The runtime under these chromatographic conditions was 10 min. The method was linear over the range of 0.02–75 μg m L − 1 . The limits of detection (LOD) and quantification (LOQ) values were 4.1 and 20 ng m L − 1 , respectively. The method was successfully extended to study the effect on GNC upon treatment with 2 N NaOH, 2N HCl, and 5% H2O2 for 2 hrs at 80°C and upon exposure to UV (1200 K lux hrs) for 72 hrs and thermal (105°C) for 5 hrs. The proposed method was further applied to the determination of GNC in pharmaceuticals, with good percent recovery. The accuracy and the precision of the method were validated on intraday and interday basis in accordance with ICH guidelines.
Item Type: | Article |
---|---|
Subjects: | C Chemical Science > Chemistry |
Divisions: | Department of > Chemistry |
Depositing User: | C Swapna Library Assistant |
Date Deposited: | 30 Aug 2019 06:17 |
Last Modified: | 02 Jul 2022 09:48 |
URI: | http://eprints.uni-mysore.ac.in/id/eprint/7364 |
Actions (login required)
View Item |